The High Resolution Camera (HRC ) is a microchannel plate (MCP )
instrument comprised of two detectors, one optimized for imaging
(HRC-I ), and one (HRC-S ) serves as a readout for the Low Energy
Transmission Grating (LETG ) discussed in Chapter 9. The
HRC-I provides the largest
field-of-view (
) of any detector aboard
Chandra , and its response extends to energies below the sensitivity of the ACIS (Chapter 6), albeit without comparable spectral resolution. The time resolution of
the HRC detectors (16
sec) is the best on the observatory, but can only be exploited under certain
conditions as discussed in Section 7.10.
A schematic of the HRC layout is shown in Figure 7.1, and a summary of the characteristics is given in Table 7.1. A cross-section of the HRC-S layout, the relationship to the optical axis and to the LETG Rowland circle is shown in Figure 7.2.
![]() |
![]() |
The HRC is a direct descendant of the Einstein (Giacconi et al. 1979) and ROSAT High Resolution Imagers (HRIs) (David et al. 1996). The ROSAT HRI had the same coating (CsI) as the HRC .
The Instrument Principal Investigator is Dr. Stephen S. Murray of the Smithsonian Astrophysical Observatory.
| Focal Plane Arrays | ||
| HRC-I : | CsI-coated MCP pair | |
| ( |
||
| HRC-S : | CsI-coated MCP pairs | 3- |
| Field of view | HRC-I : |
|
| HRC-S : | |
|
| MCP Bias angle: | |
|
| UV/Ion Shields: | ||
| HRC-I : | 5520 Å Polyimide, 763 Å Al | |
| HRC-S : | ||
| Inner segment | 2750 Å Polyimide, 307 Å Al | |
| Inner segment ``T'' | 2750 Å Polyimide, 793 Å Al | |
| Outer segment | 2090 Å Polyimide, 304 Å Al | |
| Outer segment (LESF) | 2125 Å Polyimide, 1966 Å Al | |
| Spatial resolution | FWHM | |
| HRC-I : pore size | 10 |
|
| HRC-S : pore size | 12.5 |
|
| HRC-I : pore spacing | 12.5 |
|
| HRC-S : pore spacing | 15 |
|
| pixel size (electronic readout) |
|
|
| [0.13175 arcsec pixel |
||
| Energy range: | |
|
| Spectral resolution |
|
|
| MCP Quantum efficiency | 30% @ 1.0 keV | |
| 10% @ 8.0 keV | ||
| On-Axis Effective Area: | HRC-I , @ .277 keV | |
| HRC-I , @ 1 keV | |
|
| Time resolution | 16 |
|
| Limiting Sensitivity | point source, 3 |
|
| (power law spectrum: |
||
|
|
||
| On-orbit | HRC-I | 9
|
| quiescent background | HRC-S | 1.8
|
| (prior to ground processing) | (0.07Å |
|
| Intrinsic dead time | 50 |
|
| Constraints: | telemetry limit | 184cts s-1 |
| maximum counts/observation/aimpoint | 450000 cts | |
| linearity limit (on-axis point source) | ||
| HRC-I | |
|
| HRC-S | |
|
|
|
Figure 7.3 illustrates the features of the HRC MCP
s. X-rays enter through an UV/Ion shield,
necessary in order to reduce/avoid signals from UV light, ions, and
low energy electrons. Most of these X-rays are then absorbed in the
CsI-coated walls of the first (input) of two consecutive MCP s. The
axes of the millions of tubes that comprise the input and output MCPs
are not parallel to the optical axis but are canted (``biased'') at an
angle of
, but in opposite directions as
shown. This bias is introduced to improve the probability of an
interaction. The CsI coating serves to enhance the photoemission over
that from a bare MCP . The resulting photoelectrons are then
accelerated by an applied electric field. The next interaction with
the walls releases several secondary electrons and so on until a
cascade of electrons is produced.
LINK TO POSTSCRIPT FILE FOR Figure 7.3
One purpose of the second (output) MCP is to provide additional gain. In addition, reversing the direction of the second MCP s bias angle with respect to the first removes a clear path for positive ions, and hence reduces the possibility of (positive) ion feedback - wherein an accelerated ion moving in the opposite direction as that of the electrons ends up causing the release of electrons and starts the process all over again.
The electron cloud -- typically about
electrons per
photon -- that emerges from the output MCP is accelerated towards a
position-sensitive charge detector. The HRC employs two types of
charge detectors: HRC-I uses a crossed grid charge
detector; the HRC-S uses a hybrid
where one axis is comprised of wires, while the other has gold lines
deposited on a ceramic substrate. Adjacent wires (or lines) are
resistively connected and every eighth wire is attached to a
charge-sensitive amplifier, referred to as a ``tap'', as illustrated
in Figure 7.4.
The X-ray position is determined by calculating the centroid of the
charge cloud exiting the rear MCP via the ``three tap algorithm''.
In short, the three tap algorithm determines the charge cloud centroid
using a combination of digital and analog electronics and off-line processing.
Fast discriminators and logic circuits first determine a ``coarse'' position, which
is based on the amplifier with maximum detected charge. Analog switches then select the
three amplifiers centered on that coarse position and steer them
to analog to digital converters. The coarse position and three digitized
values are then telemetered to the ground and used off-line to calculate the
event position. This process is performed for each axis.
The reconstructed X-ray position can then be written as
the sum of a coarse position and a charge centroid term centered on the coarse position:
| (7.1) |
![]() |
The three-tap position algorithm described above can
be improved upon by making use of the
predictability of the shape of the charge cloud
exiting the rear MCP.
The spatial distribution of the charge cloud
leaving the rear of the 2nd MCP has a very
specific shape for X-ray induced events.
This shape has often been modeled
as the combination of a Gaussian and
a Lorentzian distribution. Due to this specific
shape, it has been observed and simulated via
Monte Carlo techniques that the fine position
term:
| (7.2) |
| (7.3) |
For more details concerning the HRC see Murray & Chappell (1989) and Zombeck et al. (1995) and references therein.
The spacecraft is dithered during all observations in a Lissajous figure. For observations with the HRC , the dither amplitude is 40 arcsec peak-to-peak, with nominal periods of 1087 (in Y) and 768 (in Z) seconds. Dither serves the purpose of smoothing pixel-to-pixel variations in the response. The dither also eliminates gaps in coverage caused by the HRC-S inter-segment spaces near -50 A and +60 A (see Figure 7.3). The effects of dither are removed during high-level ground processing of the data.
Imaging with the HRC is best performed with the HRC-I because of the much lower background (Section 7.8) and larger field of view. The intrinsic PSF of the
HRC is well modeled by a gaussian with a FWHM of
m (
0.4 arcsec).
The HRC pixels, determined by the electronic readout and not the pore size,
are
m (0.13175 arcsec). The HRC response is thus well matched to the
intrinsic HRMA resolution (Chapter 4).
Approximately 90% of the encircled energy lies within a 14 pixel diameter region (1.8 arcsec) from the center pixel for the observation of AR Lac shown in Figure 7.5. The measured PSF is as good or better than the simulations because a very conservative pre-flight estimate of the aspect solution was used in the simulations.
The imaging resolution of the HRC-I /HRMA combination degrades off-axis for two reasons: the HRMA PSF increases with increasing off-axis angle and the deviation increases between the flat HRC-I detection surface and the curved HRMA focal surface. The off-axis imaging behavior of the HRC-I /HRMA is shown in Figure 7.6. The nominal best-focus of the HRC-I is chosen to provide the best image quality in the center of the field-of-view.
![]() |
![]() |
![]() |
The shields are useful in suppressing out-of-band (outside the X-ray band)
radiation from the ultraviolet through the visible.
The detector response to out-of-band light for an object in its field-of-view is
a possible source of unwanted signal. Suppressing out-of-band radiation is particularly
important for observing sources which have bright XUV and UV fluxes. The
HRC has strongly reduced sensitivity in this spectral region,
as shown in Figure 7.8. As part of the in-flight calibration program
the bright A star Vega (A0V, U=0.02, B=0.03, V=0.03) was observed
with both the HRC-I and HRC-S . The predicted
count rate for HRC-I was
cts s-1 . The observed upper
limit is
cts s-1 . The image of Vega was also placed on three regions of
the HRC-S - the inner segment ``T'', the thin aluminum inner segment, and on one
of the thin aluminum outer segments. The predicted count rates were 1, 400, and 2000
cts s-1 , respectively. The corresponding observed rates were 0.2, 240, and 475
cts s-1 . Sirius was observed with the HRC-S /LETGS in order to obtain a soft
X-ray spectrum of Sirius B (white dwarf) and Sirius A ( A1V, V=-1.46, B-V=0.01) was
seen in zeroth order at about the expected count rate. Based upon these sets of
observations, the UV/Ion shields are performing as designed. Other
stars will be observed as part of future calibration activities.
For a detailed discussion of the out-of-band response of the HRC to stars,
the reader is referred to
http://hea-www.harvard.edu/HRC/calib/palermopaper.ps.
Plots of the out-of-band count rate for any star for which T
, V, and N
are
known, are given.
Scattered UV, far-UV (FUV), and
extreme-UV (XUV) light from the Sun or bright Earth may cause a background strongly
dependent on viewing geometry. The spacecraft was designed to limit the contribution
from stray scattered radiation to 0.01 cts cm
s
(
cts arcsec
s
) on
the HRC . The imaged
components of scattered radiation are dependent on the solar cycle, but are
at most
0.01
for most lines of sight.
![]() |
The quantum efficiency of the HRC detector is the product of the appropriate UV/Ion shield transmission and the quantum efficiency of the CsI coated MCP . Pre-flight flat field measurements show a 10% variation in the QE across the HRC-I . The HRC-S also exhibits QE variations of the same magnitude, where the complex structure of the HRC-S UVIS contributes to the spatial variations.
The combined HRMA /HRC-I and -S effective areas - the product of the HRMA effective area, the quantum efficiency of the HRC and the transmission of the appropriate UV/Ion shield - are shown, integrated over the point spread function, in Figure 7.9. See also the discussion in the HRMA chapter (Section 4.2.2, and Figures 4.2 and 4.3.) Analysis of data from in-flight calibrations is ongoing (See http://cxc.harvard.edu/cal/Hrc/).
![]() |
The HRC-I total counting rate on orbit is about 250 cts s-1 due mostly to cosmic ray
events which are detected in the anti-coincidence shield (antico) and flagged.
The observed cosmic ray
event rate is about twice what was assumed pre-launch and corresponds to
2
. Because a rate of 250 cts s-1 saturates the telemetry
limit of 184 events sec
, the on-board veto function has been
activated. This reduces the valid event rate to about 50 c/s over the field
yielding a background telemetered rate of 10
.
The background is generally flat,
or at least smoothly varying over the field with no more than a 20% difference
between the center (higher) and edges (lower) of the detector. Note, the total event
rate remains unchanged, but detector events in coincidence with antico events no longer enter the telemetry data stream. Before
launch the expected rate, after vetoing the effects of cosmic rays, was 10-20 cts s-1 composed of
mainly the internal rate of the MCP s (10-15 cts s-1 ),
and a small contribution from cosmic rays due to antico inefficiency. There is additional
background in the HRC-I that is not well understood. However, for point source detection and exposure times of 100 ks or less the background is virtually negligible. For studies of extended objects of low surface brightness, however, even this relatively low rate can become significant depending
on the specific details of the feature.
Ground based data analysis screening techniques have been developed
to further reduce the non-X-ray background in the HRC detectors. These
techniques are described in detail in Murray et al. (2000), Juda et al.
(2000) and Wargelin et al.
(http://cxc.harvard.edu/cal/Letg/Hrc_bg/). The net result
is that for the HRC-I, non-X-ray background is reduced by
40
compared to non-screened data while the corresponding reduction in
X-ray events is reduced by less than a few percent. For the HRC-S,
non-X-ray background is decreased by
50
and the X-ray loss is
1 to 2
. Furthermore, the screening process makes the spatial distribution of
the detector background more uniform. The screening algorithms have
the added benefit of removing saturated events which were
responsible for a faint secondary ``ghost'' images (see Section 7.10).
The HRC IPI team continues to work on the screening/processing techniques and to develop new ones. These event screening procedures have been incorporated by the Data Systems Group into the standard CXO/HRC data pipeline and are transparent to the users. Proposers interested in the details of the screening procedures are encouraged to read the aforementioned papers and references contained therein.
The anti-coincidence shield of the HRC-S is not working
because of a timing error in the electronics. The error is not correctable. As a result the event rate is very high and exceeds the telemetry rate limit. To cope with
this problem the HRC Team has defined a ``spectroscopy region'' which is about 1/2 of the full width and extends along
the full length of the HRC-S detector. The spectroscopy region is
mm x 300 mm in size. The region is imposed by using the edge blanking
feature of the electronics. With this change, the telemetered quiescent background rate is about
85 cts s-1 .
As with the HRC-I , this background can be further reduced in ground data
processing by using pulse height filtering that preferentially selects X-rays
over the cosmic ray events. A reduction in background of a factor of about
three for dispersed spectra is possible. Thus there are two relevant
background rates for the HRC-S: a telemetry rate of 85 cts s
and a
post-processing rate for calculating signal to noise. The latter is
discussed in detail in Chapter 9 (see especially
Figure 9.28).
Both the HRC-I and HRC-S experience occasional fluctuations in the background due to charged particles. These times of enhanced background are typically short (a few minutes to a few tens of minutes) and are anywhere from a factor a 2 to a factor of 10 over the quiescent rates. The increased background appears to be uniformly distributed over the detector and introduces no apparent image artifacts. On the average it seems that no more than about 20% of the observing time is affected by these events, and they are easily recognized in the secondary science rate data and so can be filtered out if desired. An example of this behavior is shown in Figure 7.10. The reader is referred to Juda et al. (2002) for more information on the HRC background.
![]() |
Examples of the minimum flux detectable by the HRC-I in a 1
arcsec radius circle on-axis and for a 20
diameter extended source
are shown as a function of observing time
and spectral form in Figure 7.11.
![]() |
Initial observations with the HRC-I detector showed a faint
secondary ``ghost'' image associated with the PSF. This
``ghost'' image was a displaced, weaker (
) image
on one side of every source in the HRC-I field of
view. The cause of this
imaging anomaly was determined to be saturation of fine
position amplifiers. The previously mentioned event processing
algorithms which are now part
of the CXO/HRC data pipeline have reduced the relative intensity of
the ghost image to
effectively eliminating this anomaly. A
similar ``ghost'' image existed in the HRC-S but at a much reduced
intensity.
A problem has been discovered in HRC event times. A wiring error in the detector causes the time of an event to be associated, not with that event, but with the following event, which may or may not be telemetered. The result is an error in HRC event timing such that the accuracy is degraded from about 16 microseconds to a time which is more or less the mean time between events. To date, the total HRC-I rate has been dominated by the background, and is about 250-300 events/sec. Thus, if the time tag of each event is shifted, it turns out (verified by laboratory tests and simulations) that the average uncertainty in the time of any event is just the mean time between events. For example, if the trigger rate is 250 events/sec, then the average uncertainty in any time tag is less than 4 milliseconds.
The HRC Team has worked out a special operating mode that allows high precision timing to be achieved (see Section 7.13.1). This new timing mode uses only the central segment of the HRC-S Disabling the outer two segments lowers the total counting rate by two-thirds, dropping it below the telemetry saturation limit for most sources. Thus, there is a high probability that all events will be telemetered. In this case, once the time tag of each event has been appropriately shifted in ground processing, the original timing accuracy (16 microseconds) can be recovered. In using this approach, it is prudent to be sure that the total counting rate, source plus background, is somewhat below the telemetry saturation limit to avoid the effects of statistical fluctuations.
In addition to the primary science data for individual events,
the rate of microchannel plate triggers (total rate)
and triggers (valid rate) that pass on-board
validity tests are telemetered to the ground. The valid rate is used to correct
the primary rate for deadtime and telemetry saturation effects. As long as the
primary rate is below saturation, the primary rate itself can be used to make the
small (
1%) correction, since the event processing dead-time is known. However,
when the event rate exceeds saturation, a not uncommon occurrence because of the
flaring background from low energy protons, the valid rate is necessary to
correct the event rate. Unfortunately, the total and valid event rates are
overestimated by about 15% for normal operation of the HRC-S . The problem is caused by
an overshoot in occasional large trigger pulses. This results in double counting
in the total and valid event on-board scalers. The primary science event is not
affected, since once event processing starts with the initial trigger pulse,
a gate rejects further pulses until processing is complete. The HRC-I does not
have the overshoot problem. A scheme that corrects the HRC-S valid event rate based
upon the fraction of event pulse amplitudes that are above a given threshold
(segment dependent) has been developed but not yet implemented in the standard
processing software.
Calibration of the HRC included laboratory calibrations, a system-level ground calibration with the HRMA and HRC at the X-ray Calibration Facility (XRCF) at MSFC, and on-orbit calibration using celestial and on-board radioactive X-ray sources. The on-orbit calibration of the HRC is an on-going activity. All calibration data are, or will be, described in detail on our web site (http://cxc.harvard.edu/cal).
The user is urged to consult the WWW site and its pointers for the latest information.
In addition to the general Chandra observatory level constraints (Chapter 3), there are a handful of HRC -specific considerations and constraints that must be taken into account in planning an observation.
Both the gain and the quantum efficiency are adversely affected by the total amount of charge extracted from the MCP at the point of extraction. To minimize such effects, the high voltage on the detector is lowered during passage through the radiation belts and at times of very high particle radiation. To limit the impact from X-ray sources themselves, a 450,000 count limit distributed over the dither pattern from an on-axis source at a given aimpoint has been imposed. Users anticipating to exceed this value should so note in the comments section of the RPS form when submitting their proposal. In this case, the CXC will establish new aimpoints as necessary. Offsets in the pointing may be imposed, if necessary, in order to limit the accumulated dose to a given region of the MCP .
There are of course, other combinations and situations that can lead to telemetry saturation - numerous faint sources on the field, a too-bright extended source, etc.
It is important to be aware that avoiding telemetry saturation does not guarantee that linearity limits are not exceeded. There are only three approaches to assure oneself that the linearity limit is not exceeded:
For many observations, it is necessary only to specify the instrument, the exposure time, and the target coordinates. No other parameters need to be provided. There are, however, a number of optional parameters that one might wish to consider invoking to optimize a particular observation. The observer is encouraged to make use of the various tools provided when planning an observation, and not to forget to include the background when estimating sensitivity.
The HRC-S is normally operated in spectroscopy mode whereby signals from any of the three MCP segments can be recognized as triggers. An alternate mode of operation (timing) ties the signals from the outer segments to ground so that only signals from the center MCP generate triggers. A key distinction of this mode from using an edge-blanked region (described below) to select only the center MCP segment is that the timing mode selects events without using the on-board veto logic. This preferred method of doing high-precision timing observations reduces the active detector area thereby minimizing the total trigger rate. Provided that this rate is below telemetry saturation, all events will then be telemetered and the event time tags can be correctly assigned in ground processing (see Section 7.10).
The HRC-S , when used in this mode, provides about a 6 arcmin by 30 arcmin field of view.
It is possible to define a rectangular region, other than the default region, on either HRC-I or HRC-S from which events either inside (edge-blanking) or outside (center-blanking) the region are selected for telemetering. This would be done, for example, to prevent events from a nearby bright source from contributing to telemetry (see section 7.12.2). If a proposer wishes to define such a rectangular region, she/he should state this request in the "Remarks" field of the RPS form in order to prompt discussions with a CXC Support Scientist.
The HRC has a pair of blades that were designed so that quadrant focusing and mirror-tilt measurements could be performed on-orbit. The blades can be positioned to block the zero-order image in a grating observation. The positions of the blades can also be adjusted to block other portions of the field.
NOTE: The operation of the shutters is presently unavailable as an observing option.
The following is a list of publications, documents, web pages, and memos containing information about the High Resolution Camera (HRC ).
David, L.P., Harnden, F.R. Jr., Kearns, K.E, and Zombeck, M.V.
``The ROSAT High Resolution Imager (HRI) Calibration Report'', revised (1999).
http://hea-www.harvard.edu/rosat/rsdc_www/hricalrep.html
Fraser, G., ``X-ray Detectors in Astronomy'', 1989, Cambridge
University Press.
Giacconi, R., et al., 1979, Ap. J., 230, 540.
Murray, S.S., Chappell, J.H., Elvis, M.S., Forman, W.R., Grindlay, J.E., Harnden, F.R., Jones, C.F., Maccacaro, T., Tananbaum, H.D., Vaiana, G.S., Pounds, K.A., Fraser, G.W., and Henry, J.P., ``The AXAF High Resolution Camera (HRC ) and its use for observations of Distant Clusters of galaxies''
Astro. Lett. Comm., 26, 113-125, 1987.
Murray, S.S., et al., ``In-flight Performance of the Chandra High Resolution Camera'',
SPIE, 4012, 2000. http://hea-www.harvard.edu/HRC/calib/ssmspie2000.ps
Zombeck, M.V., Chappell, J. H , Kenter, A, Moore, R., W.,
Murray, S. S., Fraser, G.W., Serio, S.,``The High Resolution Camera (HRC )
on the Advanced X-ray Astrophysics Facility (AXAF)",
Proc. SPIE, 2518, 96, 1995.
http://hea-www.harvard.edu/hrc_art/hrc_spie/hrcspie0.html
Murray, S.S., Chappell, J.H., 1989, SPIE 1159, 460-475.
``Position Modeling for the AXAF High resolution Camera (HRC )''
Kenter, A., ``Degap as a Transformation of Probability
Distribution Problem'', 3/1/99.
http://hea-www.harvard.edu/HRC/calib/degap.ps
Murray, S.S., et al., ``Event Screening for the Chandra
X-ray Observatory High Resolution Camera (HRC)'', SPIE Proceedings,
4140, 2000.
http://hea-www.harvard.edu/HRC/calib/event_screening.ps
Juda, M., et al., ``Improving Chandra High Resolution
Camera event positions via corrections to cross-grid charge detector
signals'', SPIE Proceedings, 4140, 2000.
http://hea-www.harvard.edu/HRC/calib/spie2000_tap_correction.ps
Juda, M and Dobrzycki, A, ``HRC Deadtime and Telemetry Saturation'',
6/18/99.
http://cxc.harvard.edu/contrib/juda/memos/tlm_sat.html
Kenter, A.T., Chappell, J.H. Kobayashi,K.,Kraft,R.P., Meehan, G.R.,
Murray, S.S., Zombeck, M.V., Fraser, G.W., Pearson, J.F., Lees, J.E., Brunton, A.N. and Pearce, S.E.
Barbera, M., Collura, A., Serio, S.,
``Performance and Calibration of the AXAF High Resolution Camera I ''
SPIE 3114, 1997.
Juda, M., ``Telemetered vs. Processed Events'', memo, 12/7/01. http://cxc.harvard.edu/contrib/juda/memos/proc2valid/index.html
Juda, M., ``HRC-S Double Pulse Fraction'', memo, 6/27/02.
http://cxc.harvard.edu/contrib/juda/memos/proc2valid/pha_fraction.html
Zombeck, M. V., ``Secondary Science Rate Double Counting'', memo, 2/12/02.
http://hea-www.harvard.edu/HRC/calib/doublecount.html
http://cxc.harvard.edu/cal
(CXC calibration site)
http://hea-www.harvard.edu/HRC/calib/calib.html(HRC IPI Team calibration site)
Meehan, G, "Calibration of the HRC-I UV/Ion Shield", 10/13/99.
http://hea-www.harvard.edu/HRC/calib/hrci_cal_report.ps
Meehan, G.,"Calibration of the HRC-S UV/Ion Shields", 10/13/99.
http://hea-www.harvard.edu/HRC/calib/hrcs_cal_report.ps
Kenter, A.T., Chappell, J., Kobayashi, K., Kraft, R.P.,
Meehan, G.R., Murray, S.S., Zombeck, M.V., ``Performance and Calibration of the AXAF High Resolution Camera: I. Imaging Readout'', SPIE, 3114, 26, 1997.
http://hea-www.harvard.edu/HRC/calib/spie97_kenter.ps
Kenter, A., et al., ``In-flight Performance and Calibration of the Chandra High Resolution Camera Spectroscopic Readout (HRC-I )''
SPIE, 4012, 2000.
http://hea-www.harvard.edu/HRC/calib/hrci.spie2000.ps
Kraft, R.P., Chappell, J., Kenter, A.T., Kobayashi, K.,
Meehan, G.R., Murray, S.S., Zombeck, M.V., ``Performance and
Calibration of the AXAF High Resolution Camera: II. the Spectroscopic
Detector", SPIE, 3114, 53, 1997.
http://hea-www.harvard.edu/HRC/calib/spie97_kraft.ps
Kraft, R., et al., ``In-flight Performance and Calibration of the Chandra High Resolution Camera Spectroscopic Readout (HRC-S )''
SPIE, 4012, 2000.
http://hea-www.harvard.edu/HRC/calib/hrcs.spie2000.ps
Meehan, G.R., Murray, S.S. , Zombeck, M.V., Kraft, R.P. Kobayashi, K., Chappell, J.H., and. Kenter, A.T.,``Calibration of the UV/Ion Shields for the AXAF High Resolution Camera",
SPIE, 3114, 74, 1997.
http://hea-www.harvard.edu/HRC/calib/spie97_meehan.ps
Murray, S. S.; Chappell, J.H.; Kenter, A. T.; Kobayashi, K.; Kraft, R. P.; Meehan, G. R.; Zombeck, M. V.; Fraser, G. W.; Pearson, J. F.; Lees, J. E.; Brunton, A. N.; Pearce, S, E.; Barbera, M.; Collura, A.; Serio, S., ``AXAF High-Resolution Camera (HRC ): calibration and recalibration at XRCF and beyond'', SPIE, 3114, 11, 1997.
Juda, M., ``Time History of the HRC Background'', memo, 5/22/01.
http://cxc.harvard.edu/contrib/juda/memos/hrc_bkg/time_history.html
Juda, M., ``HRC Rates and High Solar Activity'', memo, 5/21/01.
http://cxc.harvard.edu/contrib/juda/memos/hrc_bkg/high_solar.html
Juda, M., et al., ``Characteristics of the On-Orbit Background
of the Chandra X-ray Observatory High Resolution Camera'', Proc. SPIE
4851, August 2002
http://cxc.harvard.edu/contrib/juda/memos/spie2002/spie2002.htm
http://cxc.harvard.edu/contrib/juda/memos/spie2002/spie2002.ps
McDowell, J., ``Coordinate Systems for Analysis of On-orbit Chandra Data, Paper I: Imaging'',
http://cxc.harvard.edu/contrib/jcm/ncoords.ps
Kenter, A.T., K.A. Flanagan, G. Meehan, S.S. Murray, M.V.
Zombeck, G.W. Fraser, J.F. Pearson, J.E. Lees, A.N. Brunton, and S.E.
Pearce, ``Microchannel plate testing and evaluation for the AXAF high
resolution camera (HRC )", Proc. SPIE, 2518, 356, 1995.
Zombeck, M.V., HRC-I out of band response.
http://hea-www.harvard.edu/HRC/calib/hrci_cal.html/uv_vis
Zombeck, M.V., HRC-S out of band response.
http://hea-www.harvard.edu/HRC/calib/hrcs_cal.html/uv_vis
Zombeck, M.V., et al., Vega calibration observations. http://hea-www.harvard.edu/HRC/calib/vega/vega.html
Zombeck, M.V., et al., ``The Out-of-band Responses of the
HRC on Chandra'', X-ray 2000 Proceedings, Palermo, 2000.
http://hea-www.harvard.edu/HRC/calib/palermopaper.ps
Zombeck, M. V., ``Response of the HRC to Vega'', memo, 10/28/02.
http://hea-www.harvard.edu/HRC/calib/vega/vega_trend.html
cxchelp@head-cfa.harvard.edu