The Chandra telescope consists of
4 pairs of concentric thin-walled, grazing-incidence Wolter
Type-I mirrors.
[X-ray optics are reviewed by B. Aschenbach (1985).]
The front mirror of each pair is a paraboloid (
)
and the back a hyperboloid (
). The eight mirrors were
fabricated from Zerodur glass, polished, and coated
with iridium on a binding layer of chromium.
The High Resolution Mirror Assembly (HRMA), shown schematically in Figure 4.1, contains the nested mirrors, center, forward and aft aperture plates, baffles, inner and outer cylinders, mounts, pre- and post-collimators, fiducial light transfer components, mirror support sleeves, forward and aft contamination covers, flux contamination monitors, and thermal control hardware. The outer mirror pair is number 1, and, progressing inwards, 3, 4, and 6. The original design had six mirror pairs; numbers 2 and 5 were eliminated. The pair diameters range from about 0.65 to 1.23 meters. The distance from the center of the Central Aperture Plate (CAP) separating the paraboloid and hyperboloid mirrors to the Gaussian focal point is 10.0548 meters, with each mirror pair varying slightly about this value. Note that this distance is close to, but not exactly, the focal length.. Some other characteristics are given in Table 4.1 and in section 4.5
| HRMA | |
| Optics | Wolter Type-I |
| Mirror coatings | Iridium |
| Nominal coating thickness | 330 Å |
| Mirror outer diameters (1, 3, 4, 6) | 1.23, 0.99, 0.87, 0.65 m |
| Mirror lengths ( |
84 cm |
| Total length (pre-collimator to post-collimator) | 276 cm |
| Unobscured clear aperture | 1145 cm |
| HRMA mass | 1484 kg |
| Focal length |
|
| Plate scale |
|
| PSF FWHM (with detector) | 0.5 arcsec |
| Effective area: | |
| @ 0.25 keV | 800 cm |
| @ 5.0 keV | 400 cm |
| @ 8.0 keV | 100 cm |
| Ghost-free field of view | 30 arcmin dia |
LINK TO POSTSCRIPT FILE for Figure 4.1
VanSpeybroeck (1997), Weisskopf (1997), and Zhao et al. (1997), describe the ground calibration of the HRMA and present some preliminary results. During these tests, the mirror assembly was mounted horizontally in a vacuum chamber and irradiated with X-rays from a source located at a distance of approximately 500 meters. The data taken in the XRCF (the X-Ray Calibration Facility at MSFC ) include the effective areas and image distributions as a function of incident energy and angle. The mirror performance during these tests differs from that expected in space because of gravity distortions and the finite source size and distance. Consequently, the calibration data cannot be directly compared to flight observations. The approach taken was to develop a model based upon surface and assembly measurements taken before the X-ray calibration activity. The X-ray calibration data then were used to validate this model and to make minor adjustments in model parameters to achieve satisfactory agreement with the observations. Further minor modifications were made as a result of flight experience.
The HRMA characteristics illustrated in this chapter were generated by a ray-trace program using this model. Note that this chapter typically gives characteristics of the HRMA only; unless otherwise indicated, blurring caused by the detector and the aspect solution is not included. These effects are very important for on-axis sources, and are included in the instrument chapters (Chapters 6 and 7). See also section 4.4.
The unobscured geometric clear aperture of the HRMA is
1145 cm
. The obstruction
of the HRMA clear aperture by supporting struts is less than 10%.
Since reflectivity depends on energy as well as grazing angle, the
HRMA throughput varies with X-ray energy.
Figure 4.2 shows the HRMA effective area as a function of X-ray energy, in linear scale, as well as the HRMA /ACIS and HRMA /HRC effective areas. Figure 4.3 shows the same curves with log-log scale.
Figure 4.4 shows the effects of off-axis vignetting on the effective area for various energies; the plotted results are averages over the azimuthal angle.
The above predictions are based upon the mirror model discussed previously and scaled by the HRMA ground calibration data. The agreement between the effective area predicted for the ground calibration using this model and the actual measurement, as well as the scaling function used for the on-orbit prediction, are shown in Figure 4.5.
![]() |
![]() |
![]() |
![]() |
The Chandra HRMA point-spread function (PSF ) has been simulated with numerical ray-trace calculations based upon the model previously discussed. A most useful parameter is the encircled energy fraction (the two-dimensional integral of the PSF ) as a function of radius from the image center. The PSF and the encircled energy fraction for a given radius depend upon off-axis angle and energy.The HRMA optical axis is defined for practical purposes, and calibrated in flight, as the direction of the narrowest PSF . The PSF broadens, and the encircled energy fraction decreases, as the off-axis angle increases because of mirror aberrations. The performance also degrades as the X-ray energy increases because of larger X-ray scattering.
Figure 4.6 shows the encircled energy fraction as a function of image radius for on-axis rays and for different energies. The resulting increase in image size with energy is apparent. The fraction contained within one and ten arc seconds diameters for on-axis rays is shown as a function of energy in Figure 4.7 and in Table 4.2.
![]() |
| X-ray: | Encircled Energy Fraction | ||
| Diameter | |||
| kev | Å |
|
|
| 0.1085 | 114.2712 | 0.7954 | 0.9979 |
| 0.1833 | 67.6401 | 0.7937 | 0.9955 |
| 0.2770 | 44.7597 | 0.7906 | 0.9929 |
| 0.5230 | 23.7064 | 0.7817 | 0.9871 |
| 0.9297 | 13.3359 | 0.7650 | 0.9780 |
| 1.4967 | 8.2838 | 0.7436 | 0.9739 |
| 2.0424 | 6.0706 | 0.7261 | 0.9674 |
| 2.9843 | 4.1545 | 0.6960 | 0.9560 |
| 3.4440 | 3.6000 | 0.6808 | 0.9479 |
| 4.5108 | 2.7486 | 0.6510 | 0.9319 |
| 5.4147 | 2.2898 | 0.6426 | 0.9300 |
| 6.4038 | 1.9361 | 0.6365 | 0.9344 |
| 8.0478 | 1.5406 | 0.5457 | 0.9185 |
| 8.6389 | 1.4352 | 0.5256 | 0.9151 |
| 10.0000 | 1.2398 | 0.4971 | 0.8954 |
![]() |
Pre-flight measurements and also images taken at the XRCF
show that there is a slight (
m)
offset between the optical axes of the paraboloids and hyperboloids,
and that pair 6 is slightly
tilted with respect to the other three. Consequently,
the image from mirror pair 6
is not as symmetrical as the images from the other
shells. The effect of this asymmetry on images depends on energy
because of the different relative contribution of mirror pair 6.
Figure 4.8
shows simulated HRMA /HRC-I images at four energies.
The effect of the mirror pair 6 alignment errors can be seen in
the 8.6 keV image; note that the core of this image is
not centered in the distribution.
This
effect
is not important compared to other sources of image degradation
encountered in flight such as uncertainties in the aspect solution.
![]() |
The PSF broadens for off-axis sources, and there is considerable distortion in the image even for a perfect mirror set. This distortion is due to the aberrations of Wolter type I optics and to the different focal surfaces for the four mirror pairs. The increase in image size with off-axis angle is greatest for the inner shell, and hence is larger for higher X-ray energies.
Figure 4.9 shows the dependence of encircled
energy radii on off-axis angle on the HRC-I with the HRMA focus at the
HRC-I aimpoint. Because the HRC-I is axially symmetric with respect
to the HRMA optical axis, the off-axis encircled energy radii are
almost azimuthally symmetric, except some small asymmetry due to the
imperfect HRMA as mentioned above. The figure gives the averaged
radii for 1.49 keV and 6.40 keV at 50% and 90% encircled energy.
The blurs due to the HRC-I spatial resolution and the Chandra aspect
error (FWHM: 0.22
) are included.
The ACIS-I surface is not axially symmetric with respect to the HRMA optical axis, because the HRMA aimpoint is located at the inner corner of one of the four ACIS-I chips - I3. Thus the off-axis encircled energy radii are not azimuthally symmetric. Figure 4.10 shows the dependence of encircled energy radii on off-axis angle on the four ACIS-I chips. The figure gives the encircled energy radii for 1.49 keV and 6.40 keV at 50% and 90% encircled energy in four azimuthal directions - from the aimpoint to the outer corners of the four ACIS-I chips. The blurs due to the ACIS-I spatial resolution and the Chandra aspect error are included.
![]() |
![]() |
Figure 4.11 contains simulated HRMA images
of point sources
and
off-axis, on the flat
HRC-I surface, with and without the HRC-I detector effects. At
off-axis the separation between the flat detector
surface and the curved focal surface is significant. Cusps in the
HRMA images are due to a slight misalignment of the parabolic and
hyperbolic mirrors.
![]() |
There was a concern that the effective areas might change between ground calibration and flight due to accumulation of contaminants or degradation of the reflecting surfaces. A Flux Contamination Monitor (FCM) was added to evaluate these effects. The FCM consisted of radioactive sources embedded in the forward contamination cover. The ACIS response to these sources was measured in the XRCF at the end of ground calibration, and again in orbit before the forward contamination cover was removed. No change in performance was detected.
The FCM is described in Elsner et al., 1998. The Chandra detectors are not exposed to the FCM now that the forward contamination cover has been opened.
Baffles prevent non-reflected or singly reflected rays from impinging
on the focal plane within the central
diameter of the
field of view. Outside of this region, however, singly reflected rays
from strong off-axis sources may appear. Figure 4.12
shows a simulated ghost image in which the off-axis angle was chosen
to give the closest approach of singly-reflected rays to the focal
point. In this situation, the outermost ray could extend into the
HRC field of view. The area extending from the optical axis to a
diameter of about 30
is free from ghost images.
![]() |
The HRMA performance discussed in the previous sections will be slightly degraded
by uncertainties in the aspect solution and the details of the imaging detector spatial response function. The ground software system
also deliberately adds a small random position error to reduce image artifacts
which result from instrument and data system integer location values
(these can be removed if desired.)
These effects are illustrated for the HRC-I and HRC-S instruments
in Figures 4.13 and
4.14 respectively. These figures
also show the fractional encircled energy as a function of radius
actually observed in flight compared to model calculations at 0.277
and 6.403 keV. A aspect error of 0.22
(FWHM) was
included in the model calculations. The agreement between the
observations and the lower energy model predicted curve is quite
good. The HRC efficiency and typical spectral photon number
densities both decrease with energy, so the higher energy response
curve would not be expected for most sources.
![]() |
![]() |
Similar calculations have been performed for the ACIS-S (S3) over a wider range of energies; the results are shown in Figure 4.15. The simulation accounted for the typical spacecraft jitter, so the location of the instrument pixel boundaries has little effect. There is, however, a small effect of the location of the source compared to the data system pixel boundaries. These particular calculations were performed for a point source centered on the boundary between two data system pixels. The ACIS-I instrument response is similar.
![]() |
The curves of this section, Figures 4.13, 4.14, and 4.15 may be compared with Figure 4.6 to estimate the image performance degradation due to non-HRMA effects.
The Chandra mirrors represent a logical progression from those of the
EINSTEIN (HEAO-2) (Giacconi et al. 1979) and Rosat (Trümper
1983; Aschenbach 1991) missions. Each of these
previous X-ray observatories utilized nested Wolter Type-I optics
with about 4 arcsec angular resolution. The
EINSTEIN mirror assembly had considerably less geometric area
than Chandra ,
while Rosat had comparable area (
) at low energies
(
).
The Validation Engineering Test Article-I (VETA-I)
was constructed to verify the technology required for the spatial resolution of Chandra .
VETA-I contained the
proto-flight mirror shells
constructed to final tolerances, but uncoated and with ends uncut. The
VETA-I
tests included the image full-width-half-maximum, encircled energy, effective
area, and ring focus properties
(for azimuthal and low spatial-frequency figure).
Many of the results of these tests appear in SPIE Proceedings 1742,
e.g. Hughes et al. 1992.
Zhao et al. 1993, give a good overview of the VETA tests.
A series of papers in SPIE Proceedings 3113,
e.g. Zhao et al. 1997, report on the
XRCF calibration of the HRMA assembly.
| Hyperboloid | ||||
| Mirror | [rad] | [deg] | [rad] | [deg] |
| 1 | 0.05961 | 3.42 | 0.06115 | 3.50 |
| 3 | 0.04795 | 2.75 | 0.04923 | 2.82 |
| 4 | 0.04223 | 2.42 | 0.04237 | 2.49 |
| 6 | 0.03144 | 1.80 | 0.03232 | 1.90 |
Extensive measurements of the mirror shapes and of the surface characteristics were made at Hughes-Danbury Optical Systems (HDOS) during fabrication of the mirror segments and during assembly at Eastman-Kodak Co. HRMA throughput depends critically on the coating of the individual mirror elements carried out at Optical Coating Laboratory, Santa Rosa, California. Mirror flats were present in the coating chamber and coated with iridium at the same time as the HRMA mirror elements. Reflectivity of X-rays from these witness flats was measured, and some results have been described in the 1997 July/Aug SPIE meeting proceedings.
Insulation and heaters maintain the HRMA temperature at
Fahrenheit on-orbit to minimize changes from the assembly and
alignment environments.
Postscript copies of various aspects of the HRMA calibration can be obtained from the CXC HRMA Calibration Group:
http://hea-www.harvard.edu/MST/simul/xrcf/report/index.html
Further information can be obtained from the MSFC Chandra calibration page http://wwwastro.msfc.nasa.gov/xray/xraycal/
cxchelp@head-cfa.harvard.edu